Quick Answer: A spot reflecting galvanometer is a sensitive suspended-coil galvanometer in which a small mirror on the coil reflects a light beam onto a scale, so a light spot replaces the pointer. It is used mainly as a null detector in potentiometer and bridge experiments, where it must reveal very small currents.
The Optical Lever Inside the Case
The heart of the instrument is a coil suspended in the field of a permanent magnet by a fine strip, with a tiny mirror fixed to the suspension. Current in the coil produces a torque that turns it until the twist of the suspension balances the torque. A lamp and lens send a beam onto the mirror, and the reflected beam forms a spot on a scale.
This arrangement is an optical lever. When the mirror turns through a small angle, the reflected beam turns through twice that angle, and the longer the path of the beam, the further the spot travels for the same twist. In cabinet designs the beam is often folded between internal mirrors so that a long light path fits inside a compact case. The result is a deflection far larger than a pointer of reasonable length could give for the same current.
Because the light beam has no mass, the coil can be very light and the suspension very fine, which is what gives the spot reflecting galvanometer its sensitivity. That same fineness is why the instrument must be clamped before it is moved.
Finding a Null
In a potentiometer experiment the galvanometer sits in the branch that compares two emfs. Contact is made at different points along the wire until the spot stays still when the key is pressed. Good practice is to start with a high resistance or shunt protecting the galvanometer, find the approximate balance, and then remove the protection for the final, sensitive setting. The same technique applies on a meter bridge or a Carey Foster bridge.
Specifications
| Type | Suspended-coil galvanometer with reflected light spot |
| Principle | Optical lever: beam turns through twice the angle of the mirror |
| Main use | Null detection and small-current measurement |
| Indicator | Light spot on a scale, lamp built into the case |
| Sensitivity, scale and lamp supply | Confirm at enquiry |
Applications
- Null detection in potentiometer experiments, such as comparing cell emfs or measuring internal resistance
- Balance detection in meter bridge and Carey Foster bridge experiments
- Measuring the small emf of a thermocouple
- Determining the figure of merit of a galvanometer
Care & Handling
- Clamp the suspension before moving the instrument and unclamp it only when it is level on the bench.
- Protect it with a series resistance or shunt until the circuit is close to balance.
- Keep it away from vibration, draughts and large steel objects, which can disturb the spot.
- Switch off the lamp supply when not in use, and keep the scale window clean for a sharp spot.
Why Choose LabEquip
College and senior-school laboratories that run potentiometer experiments and bridge practicals buy the spot reflecting galvanometer as their sensitive null detector. LabEquip supplies it among its physics lab equipment, and the sensitivity and supply details for your bench can be discussed through our contact page.
Frequently Asked Questions
Why does a reflecting galvanometer give a larger deflection than a pointer galvanometer?
The reflected beam turns through twice the angle of the mirror and acts as a very long, weightless pointer. The coil itself can therefore be made light and finely suspended, which increases sensitivity.
What is meant by a null detector?
A null detector shows when no current flows in a branch of a circuit. In bridge and potentiometer methods the result is found at the setting where the galvanometer shows no deflection, so sensitivity matters more than an accurate scale.
Why should I start with a protective resistance in series?
Far from balance, the current through the galvanometer can be large enough to swing the spot violently and strain the suspension. A series resistance or shunt keeps it safe until you are close to the balance point.
What is the figure of merit of a galvanometer?
It is the current needed to produce a deflection of one scale division. It is found by passing a known small current through the galvanometer and dividing by the deflection it produces.
Why must the instrument be clamped before moving?
The coil hangs from a very fine suspension strip. A jolt while unclamped can kink or break the strip, so the clamp holds the coil still during transport.
Does the spot reflecting galvanometer work on AC?
As a moving-coil instrument it responds to direct current. An alternating current reverses too quickly for the coil to follow, so the spot shows little or no steady deflection.
Last Updated: October 2026










Reviews
There are no reviews yet.