Quick Answer: The Potentiometer with Jockey Pencil is a slide-wire potentiometer: a long uniform resistance wire stretched beside a scale on a board, with a pencil jockey to find the balance point. This potentiometer measures potential differences by null comparison and is used to compare the EMFs of cells and to find a cell’s internal resistance.
Measuring Without Drawing Current
A steady current from a driver cell flows along the wire, so the potential falls uniformly with length. The cell under test is connected, through a galvanometer, from the high-potential end of the wire to the jockey. When the jockey finds the point of zero deflection, the cell’s EMF exactly equals the potential difference across that length of wire, and no current flows from the cell at all.
That null condition is the advantage over a voltmeter, which always draws some current and therefore reads the terminal voltage, slightly below the EMF. Comparing two cells needs only their two balancing lengths, E1/E2 = l1/l2. For internal resistance, the balance length is found with the cell on open circuit and again with a known resistance R across it, giving r = R(l1 – l2)/l2.
The wire length and the number of spans on the board are not stated in the listing title. A constantan rheostat in the driver circuit sets a convenient potential gradient, and the meter bridge uses the same jockey technique for resistance.
Experiments
- Comparing the EMFs of two cells, such as a Daniell and a Leclanche cell
- Finding the internal resistance of a cell
- Measuring the potential gradient along the wire
- Calibrating a voltmeter or an ammeter against the potentiometer
Specifications
| Type | Slide-wire potentiometer |
| Wire | Uniform resistance wire beside a scale on a board |
| Contact | Jockey pencil |
| Method | Null (zero-deflection) comparison with a galvanometer |
| Wire length and number of spans | Confirm at enquiry |
Care & Handling
- Use a driver cell whose EMF is greater than any EMF to be measured.
- Join the positive terminals of the driver cell and the test cell to the same end of the wire.
- Tap the jockey on the wire rather than sliding it, and keep the driver key closed only while taking readings.
- Keep the wire clean and taut, and never rest objects on it.
Why Choose LabEquip
Senior secondary and undergraduate laboratories buy this potentiometer because EMF comparison and internal resistance are standard practicals in most physics courses. LabEquip lists it in Physics Equipments, and full bench sets can be arranged through our contact page.
Frequently Asked Questions
Why is a potentiometer more accurate than a voltmeter for measuring EMF?
At the balance point no current flows from the cell under test, so there is no voltage drop across its internal resistance and the true EMF is measured. A voltmeter always draws a little current and reads slightly less.
Why can I not find a balance point?
Either the driver cell’s EMF is smaller than the EMF being measured, or the positive terminals of the two cells are not joined to the same end of the wire. Check both, and check that the rheostat in the driver circuit is not set too high.
How do I compare the EMFs of two cells?
Find the balancing length for each cell in turn without changing the driver current. The ratio of the EMFs equals the ratio of the balancing lengths. A two-way key lets the cells be switched quickly.
How is the internal resistance of a cell found?
Find the balancing length with the cell on open circuit, then connect a known resistance across it and find the new, shorter balancing length. The internal resistance equals the known resistance times the difference of the lengths, divided by the second length.
What does a longer potentiometer wire achieve?
A longer wire gives a smaller potential drop per centimetre for the same total voltage, so the balance point can be located more finely. That is why potentiometers commonly use more wire than a meter bridge, folded into several spans.
Why must the driver current stay constant?
Each balancing length means something only if the potential gradient along the wire is the same for every reading. If the driver cell runs down or the rheostat is moved between readings, the comparison is lost.