Search

In Stock

Optical Frequency Meter

Fréquencemètre optique, Real-Time High-Resolution Digital Readout Enables precise Analysis of frequency stability and signal modulation in undergraduate wave physics and harmonics research.

Integrated Signal Threshold Calibration supports the technical Application of optical triggering principles, allowing students to quantify motion in varied ambient light environments.

Ruggedized Handheld Architecture Ensures laboratory Consistency across multi-user vocational training rotations, maintaining metrological integrity despite high-cadence handling.

₹1,340.00

Quick Answer: An optical frequency meter is a frequency counter fitted with a light sensor. It counts the pulses of light reaching the sensor, whether from a flashing lamp, a flickering source or a beam interrupted by a rotating or vibrating object, and displays how many occur each second. The optical frequency meter therefore measures frequency without any contact with the moving part.

Counting Light Pulses

A photodetector in the sensor turns each change in light level into an electrical pulse. The counter opens a gate for a fixed time, counts the pulses that arrive and divides by the gate time: f = N/t. A longer gate gives a finer reading, because one extra pulse in a 10 s gate changes the result by only 0.1 Hz, whereas in a 1 s gate it changes it by 1 Hz. The cost is waiting longer for each reading, and the motion must stay steady throughout the gate.

Rotating objects are measured by making them chop a light beam. A disc with n evenly spaced holes, turning at r revolutions per second between a lamp and the sensor, produces f = nr pulses per second, so the speed follows from r = f/n. A reflective mark on a shaft does the same job when the sensor and a lamp are on the same side.

Lamp Flicker and Stroboscope Checks

Point the sensor at a filament lamp running on 50 Hz mains and the reading comes out near 100 Hz, not 50 Hz. The lamp is brightest twice in every cycle, once when the current flows one way and once when it flows the other, so its light varies at twice the supply frequency. The same idea lets students check the dial of a stroboscope: aim the sensor at the flashes and the meter shows the true flash rate.

Applications

  • Measuring the flicker frequency of mains-powered lamps
  • Checking the flash rate shown on a stroboscope tachometer
  • Finding the speed of a slotted disc or fan from the pulses it produces in a beam
  • Measuring the frequency of a vibrating blade that interrupts a light beam

Specifications

Measured quantity Frequency of variations in light intensity
Sensor Photodetector (light sensor)
Principle Pulses counted over a fixed gate time, f = N/t
Typical sources Flashing or flickering lamps, chopped beams, reflective marks
Frequency range, display and power supply Confirm at enquiry

Care & Handling

  • Shield the sensor from room lighting when measuring a weak source, because mains-lit lamps add their own flicker.
  • Keep the sensor window clean and covered when not in use.
  • Mount the sensor firmly with a clamp so it does not move during the gate time.
  • Switch the meter off after use and keep it in a dry cupboard.

Why Choose LabEquip

Physics teachers and technical colleges use an optical frequency meter wherever a frequency has to be measured without touching the moving part. LabEquip lists it in its physics lab equipment range, and a light barrier can supply the chopped beam for rotating-disc measurements. Questions about range and sensor type can go through the contact page.

Frequently Asked Questions

What does an optical frequency meter measure?

It measures how many times per second the light reaching its sensor rises and falls. That can be the flash rate of a lamp, the flicker of a mains-lit source or the rate at which a moving object interrupts a beam.

Why does a mains-powered filament lamp read about 100 Hz rather than 50 Hz?

The lamp heats up most when the current is at its peak, and the current peaks twice in each cycle, once in each direction. Its brightness therefore varies at twice the supply frequency.

How can it measure the speed of a rotating disc?

Let the disc chop a light beam, using holes or slots around its edge. If it has n slots, the meter reads n pulses per revolution, so the rotation rate is the frequency divided by n.

How does the gate time affect the reading?

The meter counts pulses for a fixed gate time and divides by that time. A longer gate gives a finer resolution, but each reading takes longer and the motion must stay steady throughout.

Can it check a stroboscope’s flash rate?

Yes. Aim the sensor at the stroboscope and the meter counts the flashes directly, giving an independent check on the rate shown on the stroboscope’s own control.

Why does room lighting upset the readings?

Mains-powered lights flicker too, and the sensor counts that flicker along with the pulses you want. Shade the sensor, aim it closely at the source or dim the room lights.

Last Updated: October 2026

Reviews

There are no reviews yet.

Write a review

Back to Top
Select your currency
INR Indian rupee
Product has been added to your cart
Compare (0)