Quick Answer: The Weight Set, 1mg-200g, Class E2 is a set of reference weights from 1 mg to 200 g, titled in accuracy class E2, covering the working range of most analytical balances. These Class E2 calibration weights are used to check analytical balances right down to milligram loads and to calibrate lower-class weights.
Milligram Weights
Below 1 g, weights are no longer small cylinders. The milligram pieces of a set like this are thin flat sheets or short lengths of wire, with an edge or end turned up for forceps. Many sets follow a shape convention so that the value can be read from the outline as well as the marking: pieces whose value begins with 5 are pentagons, those beginning with 2 are squares and those beginning with 1 are triangles, or the matching number of wire segments. The convention helps because markings on such small pieces are hard to read and easily worn.
Milligram weights are the most delicate part of the set. A breath can move them, static charge can make them cling to forceps, and a single fingerprint is a measurable fraction of their mass. They are handled with clean non-magnetic forceps, over the case or a clean surface, and never slid across the case lining.
Testing a Balance at Low Loads
For an analytical balance the critical region is often near zero, where small samples are weighed. Two checks draw on the small end of the set. Low-load linearity compares readings of several milligram and gram weights with their values. The minimum-weight test uses repeatability: a small weight is weighed repeatedly, the standard deviation of the readings is found, and from it the laboratory calculates the smallest sample mass for which that scatter stays within its required relative accuracy. Samples below this minimum weight cannot be weighed reliably on that balance, however many digits the display shows.
Applications
- Calibration and checking of analytical balances across their working range
- Minimum-weight determination for sample weighing in regulated laboratories
- Calibrating F1 and F2 milligram and gram weights by comparison
- Metrology teaching in pharmacy, chemistry and instrumentation courses
Specifications
| Nominal range | 1 mg to 200 g, as titled |
| Accuracy class | E2, as titled |
| Milligram pieces | Sheet or wire form, handled with forceps |
| Intended use | Reference weights for analytical balances and lower-class weights |
| Piece list, material, case and calibration report | Confirm at enquiry |
Care & Handling
- Open the case of Class E2 calibration weights only on a clean, draught-free bench, and handle milligram pieces with clean non-magnetic forceps.
- Never touch any piece with bare hands; use gloves or forceps for the gram pieces as well.
- Return each piece to its labelled compartment immediately after use, especially the small sheet weights.
- Record any dropped or damaged piece and have it rechecked before relying on it again.
Why Choose LabEquip
Pharmaceutical, analytical and research laboratories buy E2 sets running from milligrams to 200 g to match their analytical balances. LabEquip lists this set in its Physics Equipments range, and the Class F1 weight set, 1 mg to 200 g is available where a lower class is sufficient. Share your balance model and documentation requirements through the contact page.
Frequently Asked Questions
Why are the milligram weights different shapes?
Markings on very small pieces are hard to read, so many sets identify milligram values by shape: pentagons, squares and triangles for values beginning with 5, 2 and 1. Wire pieces use the matching number of segments.
What is the minimum weight of a balance?
It is the smallest sample mass that can be weighed while meeting a required relative accuracy. It is calculated from the repeatability of the balance at low loads, so it depends on the balance, its location and the accuracy required.
Is a 200 g top value enough for an analytical balance?
Many analytical balances have capacities around that level, so the set can cover their full range. For larger balances, a set reaching higher nominal values is needed for the span test.
How should milligram weights be picked up?
With clean, non-magnetic forceps, gripping the raised edge or end. Lift them straight up rather than sliding them, and work over the case so that a dropped piece lands somewhere safe.
Why does static matter when handling small weights?
Static charge on forceps, plastic cases or the balance can attract or repel tiny pieces, making them jump or cling. It can also exert a force on the pan that disturbs the reading. Normal room humidity and metal forceps reduce the problem.
Can E2 weights be used to calibrate other weights?
Yes. E2 weights are used to calibrate F1 and lower-class weights by comparison on a suitable balance, which is how traceability is passed down to working weight sets.