Quick Answer: The Weight Set, 1mg-200g, Class F1 is a set of weights from 1 mg to 200 g, titled in accuracy class F1, the class between E2 and F2. A Class F1 weight set is a practical working standard for teaching, pharmacy and quality-control labs that check analytical and precision balances regularly.
The Middle of the Ladder
F1 weights are intended for checking F2 weights and for use with balances of the special and high accuracy classes. In practice they are a common choice for laboratories that need regular, documented balance checks but do not run a calibration service. Their permissible errors are about three times those of E2 weights of the same value, a difference that rarely matters for routine checks of top-pan precision balances and of many analytical balances in teaching and quality-control use.
Handling rules are the same in principle as for higher classes: gloves, forceps, a clean case and time to reach room temperature. F1 tolerances simply leave a little more margin for everyday laboratory conditions.
Building a Balance-Check Routine
A weight set earns its keep when it is used to a plan. A written routine typically has three levels. A daily check before use places one weight near the usual load on the pan and compares the reading with a set limit. A periodic check, perhaps monthly, adds linearity across several weights, repeatability with one weight and an eccentricity test with the weight moved around the pan. Finally, an external calibration by a qualified service at a fixed interval confirms both the balance and the check weights.
Each check goes into a log with the date, the weight used, the reading and the initials of the person who made it. When a reading falls outside its limit, the balance is labelled out of use until it has been investigated. The F1 set supplies the known masses for every part of this routine except the external calibration.
Applications
- Daily and periodic checks of analytical and precision balances
- Calibrating F2 weights by comparison
- Balance qualification exercises in pharmacy and chemistry teaching labs
- Practical metrology courses on weighing uncertainty
Specifications
| Nominal range | 1 mg to 200 g, as titled |
| Accuracy class | F1, as titled |
| Role | Working standard for balance checks and F2 calibration |
| Handling | Gloves and forceps |
| Piece list, material, case and calibration report | Confirm at enquiry |
Care & Handling
- Use the Class F1 weight set only for checks and calibration, not as general loads in experiments.
- Handle with gloves and forceps, and let the weights reach room temperature before checks.
- Keep the case closed between uses and store it beside, not on, the balance bench to avoid vibration and knocks.
- Recalibrate at the interval in your quality plan and after any accident.
Why Choose LabEquip
Pharmacy colleges, teaching labs and quality-control departments often settle on F1 as the right balance between accuracy and practicality. LabEquip lists this set in its Physics Equipments range; for coarser routine checks the Class F2 weight set, 1 mg to 200 g is also available. Share your balance list via the contact page.
Frequently Asked Questions
Where does Class F1 sit among weight classes?
Between E2 above and F2 below. F1 weights are used to check F2 weights and with balances of the special and high accuracy classes, making them a common working standard in laboratories.
Is F1 good enough for an analytical balance?
For routine checks of many analytical balances, yes, provided the weight’s permissible error is small compared with the balance tolerance you need. For calibrating the most sensitive balances, an E2 set is the safer choice.
What should a daily balance check include?
Level and zero the balance, place one check weight near the usual load in the centre of the pan, and compare the reading with a pre-set limit. Record the result in the balance log before weighing samples.
What is an eccentricity test?
The same weight is placed at the centre and at points around the edge of the pan. A balance that reads the same at every position has no eccentric-load error; differences show that it needs adjustment.
Why log every check?
A log shows trends. A balance that slowly drifts, or a weight that changes value, is easy to spot over weeks of records, and the log is evidence that weighings were made on a checked balance.
Can students use the F1 set?
Yes, under supervision, in practicals on balance checking and weighing uncertainty. Students should wear gloves, use forceps and return each piece straight to the case, and the set should be kept apart from general class weights.
Last Updated: October 2026










Reviews
There are no reviews yet.