Quick Answer: The Precision Balance – 220g/330g – 0.001g is an electronic top-loading balance listed in 220 g and 330 g capacity versions, both reading to 1 mg. Laboratories use this 0.001g precision balance for reagent weighing and for gravimetric checks such as calibrating pipettes and burettes by weighing the water they deliver.
Calibrating Glassware by Weighing Water
Volumetric glassware is checked most directly by mass. Fill a pipette to the mark with distilled water at a known temperature, run it into a weighed, covered flask, and weigh again. Dividing the mass of water by its density at that temperature (about 0.998 g per mL at 20 °C) gives the volume actually delivered. With a 1 mg display, a 10 mL pipette’s delivery is found to about one part in ten thousand, finer than the tolerance of the pipette itself, so the balance does not limit the check.
The same method checks burettes at several points along the scale, measuring cylinders and the larger settings of adjustable micropipettes. Repeating each delivery several times shows both the systematic error, from the mean, and the scatter, from the spread of results.
Specifications
| Type | Electronic precision balance, top-loading |
| Capacity | 220 g or 330 g version |
| Readability | 0.001 g (1 mg) |
| Typical uses | Reagent weighing, gravimetric checks of volumetric glassware |
| Draft protection, pan size and interfaces | Confirm at enquiry |
220 g/330 g Compared with the 200 g/300 g Model
LabEquip also lists a 200 g/300 g milligram balance. Both read to 0.001 g; this model’s capacities are 20 g and 30 g higher, a margin that helps when a heavy flask and its water sit near the 200 g or 300 g mark. Beyond that, the choice between the two lines comes down to the features of each model.
Care & Handling
- When weighing water for glassware checks, cover the receiving vessel between readings so evaporation does not reduce the mass.
- Let the water, the glassware and the 0.001g precision balance all reach the same room temperature before starting.
- Dry the outside of vessels before placing them on the pan; drops on the pan add mass and can corrode it.
- Keep the balance level and recalibrate it if it is moved.
Applications
- Gravimetric calibration of pipettes, burettes and measuring cylinders
- Weighing reagents for college practicals and quality-control tests
- Determining the density of liquids by weighing a known volume
- Comparing the masses of small components or samples for consistency
Why Choose LabEquip
Quality-control and teaching laboratories that check their own volumetric glassware need a milligram balance with room for flasks of water. LabEquip lists this model in its physics lab products. Ask about capacity versions and quantities on the LabEquip contact page.
Frequently Asked Questions
How is a pipette calibrated with this balance?
Weigh a covered receiving vessel, deliver the pipette’s volume of distilled water into it, and weigh again. Divide the mass of water by the density of water at the measured temperature to get the delivered volume, and repeat several times to find the average and spread.
Why must the water temperature be recorded?
The density of water changes with temperature, so the conversion from mass to volume depends on it. Near room temperature the change is small, but it is large enough to matter when glassware is checked against its tolerance.
How do I find the density of a liquid by weighing?
Weigh an empty, dry container, add a known volume of the liquid with a pipette and weigh again. The mass of liquid divided by its volume gives the density; a 1 mg balance makes the mass error negligible.
How does it differ from the 200 g/300 g model?
Both read to 0.001 g. This model’s capacity versions are 220 g and 330 g, giving slightly more room for heavy vessels.
Does evaporation affect weighings of water?
Yes. At 1 mg it shows within a minute or two from an open beaker. Use a narrow-necked flask or cover the vessel, take readings promptly, and keep the balance away from warm air from other equipment.
Why do results differ between operators?
Small differences in technique, such as reading the meniscus, the drainage time of the pipette and how long the vessel stays open, change the delivered mass. Gravimetric checks are a good way to train operators, because the balance shows these effects clearly.