Quick Answer: De Sauty’s Bridge is an AC bridge circuit for comparing two capacitances: two capacitors and two resistors form its four arms, and when the bridge is balanced the ratio of the capacitances equals the inverse ratio of the resistances. De Sauty’s bridge is a standard college experiment for measuring an unknown capacitance against a known one.
The Balance Condition
The bridge is arranged like a Wheatstone bridge, with capacitors C1 and C2 in two adjacent arms and non-inductive resistances R1 and R2 in the other two. An alternating voltage, usually from an audio-frequency oscillator, is applied across one diagonal, and a detector such as a pair of headphones or an AC null detector is connected across the other. The resistances are adjusted until the detector shows a minimum.
At balance, the points at the ends of the detector are at the same potential at every instant, so the impedances satisfy (1/ωC1)/(1/ωC2) = R1/R2. The frequency cancels, leaving C1/C2 = R2/R1. With C2 a known standard capacitor, the unknown is found from two resistance readings.
Experiments
- Comparing an unknown capacitance with a standard capacitor
- Verifying the laws of capacitors in series and in parallel
- Studying how the sharpness of the null depends on the capacitor’s losses
- Introducing AC bridges before the Maxwell, Anderson and Schering bridges
Limitations and the Modified Bridge
The simple bridge assumes both capacitors are perfect, with no energy loss in their dielectric. Real capacitors, particularly those with paper or electrolytic dielectrics, behave as a capacitance with a small resistance, so the sound in the headphones never falls to complete silence. The modified form, usually attributed to Grover, adds a resistance in series with each capacitor so that the losses can be balanced as well, giving a sharp null and a measure of the dissipation factor. With good mica or air capacitors, the simple form is adequate for teaching.
Specifications
| Apparatus | AC bridge for capacitance comparison |
| Arms | Two capacitors and two non-inductive resistances |
| Balance condition | C1/C2 = R2/R1 |
| Source and detector | Audio-frequency oscillator with headphones or an AC null detector |
| Board layout, components included and accessories | Confirm at enquiry |
Care & Handling
- Use short, tidy leads and keep them apart; long loose leads add stray capacitance that shifts the balance.
- Discharge capacitors by shorting their terminals briefly before handling them after use.
- Keep the plug contacts of resistance boxes clean and firmly seated, as a loose plug spoils the null.
- Keep the oscillator volume moderate so headphones are comfortable and the minimum is easy to judge.
Why Choose LabEquip
Undergraduate electricity laboratories buy this bridge for the AC measurement experiments, often alongside resistance and capacitance boxes. LabEquip lists it in Physics Equipments; a decade capacitance box can serve as the known capacitor. Send your practical list through LabEquip enquiries.
Frequently Asked Questions
Why does the bridge need an AC supply?
A capacitor blocks steady direct current, so with DC no current would flow through the capacitor arms once they had charged. An alternating supply keeps charging and discharging the capacitors, letting their reactance be compared with the resistances.
Why is the balance independent of frequency?
Both capacitor arms have reactances that change with frequency in the same proportion. When the ratio of reactances is set equal to the ratio of resistances, the frequency cancels out of the balance equation.
Why can I not get complete silence in the headphones?
The capacitors are not perfect: losses in their dielectric add a small in-phase component that the simple bridge cannot balance. Harmonics in the oscillator signal and stray capacitance also contribute. A modified bridge with series resistances corrects for the losses.
What frequency should the oscillator be set to?
A frequency where the ear is most sensitive, around a kilohertz, makes the minimum easiest to hear with headphones. The balance itself does not depend on frequency, so the exact value is not critical.
Can the bridge measure very small capacitances?
Small capacitances have high reactance, so the bridge becomes less sensitive and stray capacitance from the leads becomes significant. Keep leads short and use a screened arrangement for capacitances much smaller than the standard.
How do I verify the series and parallel laws?
Measure two capacitors separately, then connect them in series and in parallel and measure each combination. Compare the results with 1/C = 1/C1 + 1/C2 for series and C = C1 + C2 for parallel.
Last Updated: October 2026










Reviews