Quick Answer: The Semi Micro Balance 50/220g – 0.01/0.1mg is an electronic laboratory balance listed with two ranges: 0.01 mg readability up to 50 g and 0.1 mg readability up to 220 g. The semi micro balance is used for very small samples, such as reference standards, filters and micro-scale analyses, where the fifth decimal place of a gram matters.
Two Ranges in One Listing
The title pairs 50 g with 0.01 mg and 220 g with 0.1 mg, the usual way of describing a dual-range semi-micro balance: the fine range shows five decimal places of a gram (0.00001 g) for loads up to 50 g, and above that the display steps to four decimal places up to the 220 g maximum. That arrangement lets one instrument handle both a few milligrams of a reference standard in a light vessel and an ordinary analytical weighing in a heavier one.
Why does the fifth decimal matter? When a sample weighs only 5 mg, a 0.1 mg display limits the result to about two percent of the sample; at 0.01 mg the same sample is read to about 0.2 percent.
Specifications
| Type | Electronic semi-micro balance |
| Fine range | 0.01 mg readability up to 50 g |
| Upper range | 0.1 mg readability up to 220 g |
| Typical samples | Reference standards, filters, micro-scale analytical samples |
| Range switching, pan size and interfaces | Confirm at enquiry |
Weighing Technique at the Fifth Decimal
At this level, effects that pass unnoticed on an ordinary balance become part of the result. Breath near an open draft shield, the warmth of a hand, static on a dry plastic vial and the buoyancy of air on low-density samples can each shift the reading by several hundredths of a milligram. Users therefore let samples and containers adjust to the balance room, handle them with forceps, use small glass or metal weighing vessels, keep the shield closed while reading, and stand the balance on a stone or anti-vibration table. Filters for air-quality work are normally conditioned at a fixed temperature and humidity before both the blank and the loaded weighing.
Applications
- Weighing milligram quantities of reference standards and active ingredients
- Gravimetric measurement of particulate matter collected on filters
- Micro-scale analytical chemistry and research where samples are scarce
- Routine analytical weighing of heavier samples in the 0.1 mg range
Care & Handling
- Install the semi micro balance on a vibration-damping weighing table in a room with steady temperature, away from vents and walkways.
- Use forceps for small containers and filters; never place them on the pan by hand.
- Brush out the weighing chamber regularly, since tiny particles left on the pan affect readings at this resolution.
- Allow a full warm-up after the balance has been switched off, and recalibrate after moving it.
Why Choose LabEquip
Semi-micro balances are bought by research groups, pharmaceutical laboratories and environmental testing units whose samples are too small for an ordinary analytical balance. LabEquip lists this model in its physics equipment category; for routine 0.1 mg work a simpler analytical balance may be enough. Discuss your sample sizes and room conditions with us through the contact page.
Frequently Asked Questions
When do I need a semi micro balance instead of an analytical balance?
When samples are so small that a 0.1 mg display gives too coarse a result, typically sample masses of a few milligrams to a few tens of milligrams. The extra decimal place reduces the relative rounding error by a factor of ten.
What does dual range mean?
The balance shows a finer readability over the lower part of its capacity and a coarser one above it. This listing gives 0.01 mg up to 50 g and 0.1 mg up to 220 g.
What is air buoyancy and does it matter?
Air pushes up on everything weighed in it, by an amount equal to the weight of air displaced. Balances are calibrated with dense metal weights, so low-density samples such as powders read slightly light. In most work the effect cancels out, but in the most exacting weighings a buoyancy correction is applied.
How long should samples adjust before weighing?
Long enough to reach the temperature of the balance room: several minutes for small items and an hour or more for vessels brought from an oven or refrigerator. Temperature differences cause convection currents that make readings drift.
Why use glass or metal containers instead of plastic?
Plastics hold static charge readily, especially in dry air, and the charge exerts a force on the balance that shows as drift. Glass and metal are less prone to this, and an ionizer can neutralise charge on items that must be plastic.
Does the balance need a special table?
A heavy, rigid weighing table or a stone slab on isolating feet is strongly advised. Footsteps, nearby centrifuges and building vibration show up clearly at 0.01 mg, and an ordinary bench may flex when someone leans on it.