Quick Answer: The Nicholson Hydrometer is a floating instrument with a pan above and a weighted cup below, sunk to a fixed mark on its stem by adding small masses. It is used to find the relative density of a small solid, and of a liquid, from the masses needed to bring it to that mark.
A Constant-Immersion Method
The Nicholson hydrometer is a hollow metal float carrying a thin stem with a fixed reference mark, a small pan at the top and a weighted cup or bucket hanging below. Because it is always sunk to the same mark, the volume of liquid it displaces is fixed. Changes in the load needed to reach the mark therefore measure changes in upthrust, which is why it is described as a constant-volume, variable-load hydrometer.
For a solid, three loadings are found in water. First, W₁ is the mass on the upper pan that sinks the empty instrument to the mark. Next the solid is put on the upper pan and the mass needed falls to W₂, so the solid’s mass is W₁ − W₂. Finally the solid is moved to the lower cup, under water, and the mass needed rises to W₃ because of the upthrust on it. The relative density is (W₁ − W₂) / (W₃ − W₂).
For a liquid, the instrument’s own mass W is found on a balance, and the masses w₁ and w₂ that sink it to the mark in water and in the liquid are recorded. Since the displaced volume is the same, the relative density of the liquid is (W + w₂)/(W + w₁). The method suits small solids lighter than the load W₁. The instrument’s material and size are not given in the title; a weight box of small masses is the usual companion.
Specifications
| Type | Constant-immersion (variable-load) hydrometer |
| Parts | Hollow float, stem with fixed mark, upper pan, lower weighted cup |
| Measures | Relative density of small solids and of liquids |
| Used with | Tall jar of water or test liquid and a set of small masses |
| Material, size and accessories | Confirm at enquiry |
Experiments
- Finding the relative density of a small metal or glass specimen
- Finding the relative density of a liquid such as brine or spirit
- Teaching Archimedes’ principle through a measurement rather than a demonstration
- Comparing results with those from a hydrostatic balance or density bottle
Care & Handling
- Float the hydrometer in a tall jar wide enough that it does not touch the sides.
- Remove air bubbles from the lower cup and float before deciding that the mark has been reached.
- Add the smallest masses last and let the instrument settle between additions.
- Rinse with clean water after use in salt solutions and dry before storing.
- Handle the stem gently; a bent stem shifts the reference mark.
Why Choose LabEquip
Colleges and senior schools keep a Nicholson hydrometer for relative density practicals because one instrument handles both solids and liquids. It is in our Physics Equipments; contact LabEquip for its size and the masses it needs.
Frequently Asked Questions
How is the Nicholson hydrometer different from a common hydrometer?
A common hydrometer floats at different depths and is read from a scale on its stem. The Nicholson type is always sunk to one fixed mark, and the result comes from the masses added to reach it.
Why is the solid placed in the lower cup for the third reading?
In the lower cup the solid is under water and receives an upthrust equal to the weight of water it displaces. The extra mass then needed on the upper pan measures that upthrust, which gives the solid’s volume in terms of water.
What is the formula for the relative density of a solid?
Relative density equals the mass of the solid in air divided by its loss of mass in water: (W₁ − W₂) divided by (W₃ − W₂), using the three loadings found with the instrument in water.
Can it measure solids that float?
Not directly. A solid less dense than water would float off the lower cup. Such solids need a sinker or a different method, such as a density bottle or a hydrostatic balance with a sinker.
What limits the size of solid that can be tested?
The solid’s mass must be less than the load that sinks the empty hydrometer to the mark, otherwise the instrument sinks below the mark even with no masses on the pan.
Why must air bubbles be removed?
A bubble clinging to the float or cup adds buoyancy, so fewer masses are needed to reach the mark and the readings are wrong. Tap the instrument gently or wet it fully before taking each reading.