Quick Answer: The Newton’s Ring Experiment Microscope is a travelling microscope, a low-power measuring microscope that slides along a scale with a vernier, set up to look down on the Newton’s rings apparatus. Students set its cross-wire on either side of each interference ring and read the scale, and the ring diameters give the wavelength of light or the radius of curvature of the lens.
What the Microscope Measures in Newton’s Rings
Newton’s rings appear when a plano-convex lens of large radius rests on a flat glass plate. The film of air trapped between them is zero at the point of contact and grows thicker outwards, and light reflected from the top and bottom of that film interferes. In monochromatic light, normally from a sodium lamp, the result is a dark central spot surrounded by alternate bright and dark circles that crowd closer together towards the edge.
The microscope looks vertically down on this pattern through a plain glass plate tilted at 45 degrees, which turns the horizontal beam from the lamp down onto the lens and lets the reflected light rise into the objective. With the rings in focus, the cross-wire is set tangentially on the left edge of a chosen dark ring and the scale and vernier are read; the microscope is then moved across the centre to the same ring on the right. The difference between the two readings is that ring’s diameter.
The square of the diameter of the nth dark ring is proportional to n, so readings are taken for a run of rings and the wavelength follows from the difference of squares: λ = (D² of ring n+p − D² of ring n) / 4pR, where R is the radius of curvature of the lens surface, measured separately with a spherometer. The lens and flat plate for the experiment are in the Newton’s ring lens set.
Avoiding Backlash While Reading
The carriage of a travelling microscope is driven by a screw, and any screw has a little slack. If the screw is turned back and forth during a set of readings, that slack appears as an error. Start beyond the outermost ring to be measured, then wind steadily in one direction, reading each ring on the way across and each matching ring on the far side. Before starting, turn the eyepiece so one cross-wire is square to the direction of travel.
Specifications
| Type | Travelling (measuring) microscope for the Newton’s rings experiment |
| Viewing direction | Vertical, looking down on the lens and flat plate |
| Reading | Main scale with vernier, set by an eyepiece cross-wire |
| Used with | Newton’s ring lens set, 45-degree glass plate, sodium lamp, spherometer |
| Magnification, travel and least count | Confirm at enquiry |
Experiments and Uses
- Determining the wavelength of sodium light by Newton’s rings
- Finding the radius of curvature of a plano-convex lens when the wavelength is known
- Finding the refractive index of a liquid placed between the lens and the plate, from the shrinking of the rings
Care & Handling
- Move the carriage only with its screw; pushing the tube by hand strains the slides.
- Clean the objective and eyepiece with lens tissue, never with a handkerchief or paper towel.
- Keep grit off the scale, vernier and sliding surfaces, and fit the dust cover after each session.
- Store the microscope with clamps slackened and the tube racked to a safe position.
Why Choose LabEquip
Undergraduate physics laboratories and senior secondary schools that run the Newton’s rings practical buy this microscope together with the lens set and a sodium lamp. LabEquip lists it with its physics lab equipment, and departments fitting out several Newton’s rings stations at once can arrange the order through our contact page.
Frequently Asked Questions
Why is the centre of Newton’s rings dark?
At the point of contact the air film is almost zero thickness, but the light reflected from the glass plate below undergoes a phase change of half a wavelength. The two reflected waves therefore cancel, leaving a dark spot in reflected light.
Why are diameters measured rather than radii?
The exact centre of the ring system is hard to locate with the cross-wire, but both edges of a ring are sharp. Measuring from one side of a ring to the other is more accurate than trying to set on the centre.
Why is the result calculated from a difference of squared diameters?
The difference between the squared diameters of two rings p apart equals 4p times the wavelength times R, whatever ring is counted as the first. Dust or strain at the point of contact, which distorts the inner rings, therefore does not affect the result.
Can white light be used for Newton’s rings?
Only a few coloured rings near the centre are visible in white light, because each wavelength makes rings of a different size and they soon overlap. A sodium lamp gives many sharp rings that can be counted and measured.
How is the radius of curvature of the lens found?
Measure it with a spherometer on the curved face of the lens, or use this microscope with a light source of known wavelength and rearrange the formula to calculate R.
How does this differ from a general vernier microscope?
Both are travelling microscopes read with a vernier. This one is arranged to look straight down onto the rings; a general vernier microscope is often also used horizontally, for example on capillary bores. Confirm the mounting options of this model when you enquire.
Last Updated: October 2026










Reviews