Constant Deviation Spectrometer

Constant-Deviation Prism Assembly: Maintains a fixed telescope position while allowing precise spectral line observation.

Fine-Graduation Circular Scale with Vernier: Enables high-resolution angular measurement for accurate wavelength determination.

₹1,205.00

Quick Answer: The Constant Deviation Spectrometer is an optical instrument in which the collimator and telescope stay fixed, usually at right angles, while a constant-deviation prism is rotated to bring each spectral line onto the crosswire. A calibrated drum turning the prism lets the constant deviation spectrometer give wavelengths from a single reading once it has been calibrated.

How Constant Deviation Works

In an ordinary prism spectrometer the telescope swings round to follow each colour, and every reading needs an angle measurement and a calculation. A constant-deviation prism, most often of the Pellin–Broca type, removes that step. Light from the collimator enters the prism, is reflected internally at one face, and leaves turned through a fixed angle, usually 90 degrees, but only for the wavelength that passes through the prism at minimum deviation.

Rotating the prism changes which wavelength meets that condition, so the spectrum moves across the fixed telescope’s field. The rotation is made by a micrometer drum, whose reading therefore corresponds to the wavelength sitting on the crosswire. The relationship is not linear, so the drum is calibrated against lines of known wavelength and the readings converted with a calibration curve or table.

Experiments

  • Calibrating the drum with a mercury or helium discharge lamp
  • Measuring wavelengths of lines in an unknown emission spectrum
  • Studying the hydrogen Balmer lines and estimating the Rydberg constant
  • Observing absorption spectra of coloured solutions and filters
  • Comparing the constant deviation method with a conventional prism spectrometer

Calibrating and Measuring

Calibration uses a discharge lamp whose lines are well known, such as mercury, sodium, helium or neon. The slit is narrowed until the lines are sharp, each line is brought to the crosswire and the drum reading noted, and a graph of wavelength against drum reading is drawn. Unknown lines from another source are then measured by reading the drum and taking the wavelength from the graph. The same instrument records absorption bands when a coloured solution is placed in front of the slit.

Specifications

Instrument Constant deviation spectrometer
Dispersing element Constant-deviation prism on a rotating table
Optical layout Fixed collimator with adjustable slit and fixed viewing telescope
Reading Drum reading converted to wavelength by calibration
Light sources, drum scale and accessories supplied Confirm at enquiry

Care & Handling

  • Never touch the prism faces or lens surfaces; clean them only with lens tissue if dust appears.
  • Turn the drum gently and only within its travel; forcing it upsets the calibration.
  • Recheck the calibration with a known lamp after the instrument has been moved or knocked.
  • Discharge lamps run hot and need their own power units: let them cool before handling, and keep the dust cover on the instrument when not in use.

Why Choose LabEquip

Constant deviation spectrometers are bought by universities and colleges for atomic physics and spectroscopy practicals, typically one or two per laboratory. LabEquip lists this instrument under Physics Equipments; for the conventional prism method, see the optical spectrometer. Discuss lamps and accessories with us through the contact page.

Frequently Asked Questions

What is the advantage of constant deviation?

The telescope does not move, so wavelengths are read straight from the drum after calibration. That makes measurements quicker and avoids the repeated angle readings and calculations needed with a rotating-telescope prism spectrometer.

Why must the instrument be calibrated?

The prism’s dispersion varies with wavelength, so equal turns of the drum do not correspond to equal changes in wavelength. A calibration curve from known lines lets each drum reading be converted accurately.

Which lamp is used for calibration?

Any discharge lamp with well-known lines spread across the visible spectrum. Mercury is common because its violet, blue, green and yellow lines are bright and widely tabulated, and helium and neon add more points.

Can it measure absorption spectra?

Yes. Placing a coloured solution or filter between a white-light source and the slit shows dark bands where light is absorbed, and their positions are read on the drum and converted to wavelength.

How narrow should the slit be?

As narrow as still gives a bright enough image. A narrow slit gives sharp lines that can be set accurately on the crosswire, while a wide slit makes lines broad and lets close pairs, such as the sodium doublet, merge.

Is the Pellin–Broca prism the only constant-deviation type?

No. The Pellin–Broca is the most common, but Abbe prisms also give constant deviation, and some instruments use other designs. All work by passing light at minimum deviation for the wavelength being viewed.

Last Updated: October 2026