Quick Answer: Disposable ESR tubes are single-use graduated tubes, supplied alone or as a kit, used to measure the erythrocyte sedimentation rate: how far red blood cells settle in anticoagulated blood in one hour. They replace reusable glass ESR pipettes and remove the need to wash and dry tubes between patients.
How the ESR Test Works
In the Westergren method, venous blood is mixed with sodium citrate anticoagulant, traditionally four parts blood to one part citrate, and drawn into a vertical graduated tube. After exactly 60 minutes the height of clear plasma above the settled red cells is read in millimetres and reported in mm per hour. A raised ESR is a non-specific sign of inflammation, infection or other disease and is interpreted alongside other findings.
Disposable systems usually pair a tube that already contains citrate with a graduated pipette or column that stands in a rack, so the blood-to-citrate ratio is set by the fill mark. The reusable glass option is the E.S.R. tube.
Accurate readings depend on technique: the tube must stand exactly vertical, away from vibration, direct sunlight and draughts, at room temperature, and it must be read at 60 minutes. Tilting the tube even slightly speeds sedimentation and gives a falsely high result.
Specifications
| Test | Erythrocyte sedimentation rate (ESR) |
| Type | Single-use tubes or kit |
| Anticoagulant | Sodium citrate, as used in the Westergren method |
| Reading | Millimetres of clear plasma after 60 minutes |
Applications
- Routine ESR testing in hospital and clinic laboratories
- Haematology practicals for medical laboratory technology students
- Health camps and small labs without facilities to wash glass ESR pipettes
- Following the course of inflammatory conditions alongside other tests
Care & Handling
- Mix blood and citrate gently by inversion; do not shake.
- Stand tubes in a level rack out of sunlight, heat and vibration.
- Start a timer at filling and read at exactly 60 minutes.
- Dispose of used tubes as biomedical waste according to local rules, and never reuse them.
Why Choose LabEquip
Pathology labs, diagnostic centres, health camps and paramedical colleges use disposable ESR tubes to save washing time and reduce blood exposure. LabEquip lists the kit in the Biology Lab Products range; send your test volume and preferred format through the contact page.
Related Products
- Tube, E.S.R.: The reusable E.S.R. tube, read on the same one-hour principle in a vertical rack.
- Synthetic blood: Synthetic blood, a simulated fluid for running blood-related teaching practicals without human or animal blood.
- Pipette, Pasteur, Disposable: Disposable Pasteur pipettes for transferring small volumes of liquid during sample preparation.
Frequently Asked Questions
What does an ESR test measure?
The erythrocyte sedimentation rate measures how far red blood cells settle in a column of anticoagulated blood in one hour, read as millimetres of clear plasma. A raised ESR is a non-specific sign of inflammation, infection or other disease and is always read alongside other findings.
How long does an ESR test take to read?
An ESR is read at exactly 60 minutes after the tube is set upright. The height of clear plasma above the settled red cells is recorded in millimetres and reported as mm per hour. Reading early or late changes the result.
Why use disposable ESR tubes instead of reusable glass tubes?
Disposable ESR tubes are used once and discarded, so there is no washing, drying or risk of carry-over between samples. Many disposable systems combine a tube with sodium citrate already in it and a graduated column, so the blood-to-citrate ratio is set by the fill mark.
What can make an ESR result falsely high or low?
Tilting the tube speeds settling and gives a falsely high result; vibration, direct sunlight and a warm room also raise it. A cold room, delayed testing, clotted samples or too much anticoagulant can lower it. Fill to the mark, mix gently and read on time.
What blood-to-citrate ratio is used in the Westergren ESR method?
In the Westergren method, venous blood is mixed with sodium citrate anticoagulant, traditionally four parts blood to one part citrate. In disposable kits the citrate is often pre-filled, so the correct ratio depends on filling exactly to the marked line.
How should a disposable ESR tube stand while the test runs?
Stand the tube exactly vertical in its rack on a level bench, away from vibration, draughts, direct sunlight and heat sources, at room temperature. Leave it untouched until the 60-minute reading, then read the plasma column at eye level.