Quick Answer: Plastic cuvettes are small, square-section sample cells moulded from clear plastic, used to hold a solution in the light path of a colorimeter or spectrophotometer. They are inexpensive, usually disposable, and suit readings in the visible range.
Clear Faces, Ribbed Sides
The listing image shows two cuvettes of the familiar square design: two opposite walls are smooth and clear, and the other two are ribbed. The clear faces are the optical windows through which the light beam passes. The ribbed sides give a grip that keeps fingerprints off the windows and show which way round the cell must go into the holder.
Most disposable cells are polystyrene, which transmits well across roughly 340 to 800 nm, or PMMA (acrylic), which reaches a little further into the near ultraviolet. Below about 300 nm neither is suitable and quartz cells are used. Standard cells have a 10 mm internal path length, but confirm that this matches your instrument and method.
Both plastics are attacked by many organic solvents; acetone, for example, dissolves polystyrene. They are intended for aqueous solutions, as in most school and college colorimetry. Filling with a Pasteur pipette avoids splashing the clear faces.
Specifications
| Item | Square-section cuvette with two clear optical faces |
| Material | Clear plastic (commonly polystyrene or PMMA); type to be confirmed |
| Useful range | Visible light; not for deep ultraviolet work |
| Use | Usually disposable, single use |
Applications
- Measuring the absorbance of copper(II) sulfate standards to plot a calibration graph
- Following the rate of a reaction by colour change, such as the fading of crystal violet in alkali
- Estimating protein by the biuret method, or glucose with colorimetric reagents, in biology labs
- Measuring the turbidity of bacterial cultures at 600 nm
Care & Handling
- Hold cuvettes by the ribbed sides only, and wipe the clear faces with lint-free tissue before each reading.
- Fill to about three-quarters so the light beam passes through liquid, not the meniscus.
- Use the same cell, or a matched one, for the blank and the samples in a series.
- Discard scratched cuvettes, since scratches scatter light and raise the reading.
Why Choose LabEquip
Colorimetry now features in school and college chemistry and biology, and disposable cells spare technicians from cleaning quartz. LabEquip lists these in its chemistry lab equipment range; send your instrument details and quantity through the contact page.
Related Products
- Pipette, Pasteur, Disposable: Disposable Pasteur pipettes for transferring sample into the cuvette without splashing the clear faces.
- Wash Bottle, Plastic: A squeeze wash bottle for rinsing cuvettes with distilled water between samples.
- Cylinder, Nessler: A Nessler cylinder for visual colour comparison when no colorimeter is available.
- Fast Releaser Pipette Pump: A thumbwheel pipette pump for measuring reagent volumes accurately when preparing colorimetry standards.
Frequently Asked Questions
Which way round should a plastic cuvette go into a colorimeter?
Put it in so that the two smooth, clear faces lie in the light path and the ribbed sides face the sides of the holder. Hold the cuvette by the ribbed sides so fingerprints never reach the clear windows, and insert it the same way each time, since a small difference between the two windows can change the reading slightly.
Can plastic cuvettes be used for ultraviolet measurements?
Only in the near ultraviolet. Disposable polystyrene cuvettes transmit well across roughly 340 to 800 nm and PMMA (acrylic) cells reach a little further into the UV, but neither suits measurements below about 300 nm, such as DNA at 260 nm. Those need quartz cells or special UV-grade plastic cuvettes.
Which solvents should not be put in plastic cuvettes?
Polystyrene and acrylic cuvettes are made for aqueous solutions. Acetone, many chlorinated solvents and other strong organic solvents soften, cloud or dissolve them, and even ethanol can craze some grades over time. For organic solvents use glass or quartz cells, which also suit higher-accuracy work.
How full should a cuvette be filled?
Fill it to about three-quarters of its height so the light beam passes entirely through the liquid, well below the meniscus. Overfilling risks spills inside the instrument. Tap the cuvette gently to release air bubbles, which scatter light, and wipe the outside of the clear faces with a lint-free tissue before inserting it.
What is a blank cuvette and why is it needed?
A blank is a cuvette filled with the solvent or reagent mixture without the coloured substance being measured. The instrument is set to zero absorbance, or 100 percent transmission, with the blank in place, so that light absorbed by the plastic, the solvent and the reagents is not counted as part of the sample’s reading.
Can disposable plastic cuvettes be washed and used again?
For school colorimetry with aqueous solutions they can be rinsed with distilled water and reused a few times while the windows stay clear and unscratched. For accurate or research measurements they are used once, because scratches, residues and slight differences between cells add error. Discard any cuvette with a cloudy or marked window.