Search

In Stock

Balance, Analytical, Mechanical Type

Synthetic Sapphire Knife-Edge Bearings: Engineered with high-density synthetic sapphire to minimize friction at the fulcrum, ensuring exceptional beam sensitivity and long-term repeatability of 0.1mg resolution across the full weighing range.

Air-Damped Symmetric Beam System: Features a precision-calibrated aluminum alloy beam with integrated air damping, allowing for rapid stabilization and reduced oscillation without the need for external power or electronic transducers.

₹1,240.00

Quick Answer: A mechanical analytical balance is a beam balance of the highest sensitivity class, with knife-edges and a glazed draught case, that measures mass by balancing it against standard weights without any electrical power. It is used in chemistry and pharmacy teaching labs for quantitative weighing and to show how fine weighing works.

Two Mechanical Designs

Two mechanical patterns have been made. The older equal-arm form has two pans, a long beam on agate knife-edges and a rider or chain device for the smallest increments, so the operator balances the load against weights placed on the opposite pan. The later single-pan form works by substitution: built-in weights hang on the same side as the pan, and a sample on the pan is compensated by lifting off weights with dials until the beam returns to its reference position, with an optical scale giving the final digits.

Substitution keeps the beam load constant, so the sensitivity is the same for every weighing and unequal arm lengths do not matter. The equal-arm form is simpler to understand and makes the principle of moments visible. The title does not say which pattern this listing is, so the design and the weight-loading mechanism should be confirmed with LabEquip.

Fine Weighing Technique

The balance stands on a solid bench away from doors, windows and heaters and is levelled before use. Objects are handled with tongs or gloves because a fingerprint has a measurable mass, and they are weighed at room temperature with the side doors closed. The beam is always arrested, or put in its partial-release position, before anything is added or removed, which protects the agate edges.

For the highest accuracy, analysts allow for air buoyancy: a sample less dense than the brass or steel weights displaces more air than they do and so appears slightly light. The correction is negligible for most teaching work but matters when low-density samples are weighed precisely.

Where It Is Used

  • Gravimetric determinations in which a precipitate is dried and weighed in a crucible
  • Preparing primary standard solutions, for example from sodium carbonate
  • Teaching how balances work before students move to electronic instruments
  • Weighing where mains power is unavailable or unreliable

Specifications

Type Mechanical analytical balance
Power None required
Weighing chamber Glazed draught case with doors
Bearings Hard knife-edges on flat planes
Pattern, readability and weight loading Confirm at enquiry

Care & Handling

  • Arrest the beam fully at the end of each session and before moving the instrument.
  • Clean the pan and the floor of the case with a soft brush only; never oil the knife-edges.
  • Keep a desiccant inside the case and renew it when it is spent.
  • Check the balance against known weights at regular intervals and log the results.

Why Choose LabEquip

Colleges of chemistry and pharmacy often keep a mechanical analytical instrument for teaching alongside electronic ones, because students can see the beam, weights and rider doing the work. LabEquip offers it within Physics Equipments and also lists an analytical balance page for comparison; ask about the pattern you need through LabEquip’s contact form.

Frequently Asked Questions

Why choose a mechanical model instead of an electronic one?

A mechanical balance needs no power, has no electronic warm-up drift, and lets students see the weighing principle directly. Electronic balances are quicker for routine work, so many laboratories keep both.

What is a substitution balance?

It is a single-pan mechanical balance in which built-in weights are removed to compensate for the sample, keeping the total load on the beam constant. Because the load never changes, the sensitivity stays the same at every reading.

Why are the knife-edges usually made of agate?

Agate is very hard, can be ground to a fine straight edge and does not corrode. A sharp edge on a flat plane gives very low friction, which is what lets tiny mass differences move the beam.

What is air buoyancy correction?

Air pushes up on any object with a force equal to the weight of air it displaces. A sample less dense than the reference weights is pushed up more than they are, so it appears slightly lighter, and a calculated correction restores the true mass.

How should samples be prepared before weighing?

Bring them to room temperature, hold containers with tongs or gloves, and keep them covered if they take up moisture. Warm or fingerprinted objects give readings that drift or are slightly wrong.

Does a mechanical analytical balance need calibration?

It should be checked periodically against reference weights of a suitable class, and its rest point and sensitivity adjusted if they have shifted. A log of these checks shows when servicing is due.

Last Updated: October 2026

Reviews

There are no reviews yet.

Write a review

Back to Top
Select your currency
INR Indian rupee
Product has been added to your cart
Compare (0)