Quick Answer: The e/m by Short Solenoid Method (Magnetron Methods) apparatus finds the specific charge of the electron by placing a cylindrical diode valve inside a short solenoid and increasing the axial magnetic field until electrons can no longer reach the anode. In the magnetron method, the critical field at which the anode current falls away gives e/m.
How the Magnetron Valve Works
The heart of the apparatus is a diode valve with a straight filament cathode running along the axis of a cylindrical anode. With no magnetic field, electrons emitted from the hot cathode travel straight out to the anode and a steady anode current flows. The valve is mounted at the centre of a short solenoid, whose field runs parallel to the cathode. As the solenoid current increases, that axial field bends the electrons’ outward paths into curves.
At a particular critical field Bc the curved paths just graze the anode and turn back towards the cathode, so the anode current drops sharply. For a thin cathode and an anode of radius a, with anode voltage V, theory gives e/m = 8V/(Bc²a²). This arrangement of coaxial electrodes in an axial field is the basic geometry of the magnetron, which is where the method takes its name.
Field of a Short Solenoid and the Cut-Off Curve
Because the solenoid is short, the simple long-solenoid formula is not accurate. The field at its centre is found from the number of turns, the current and both the length and diameter of the winding: B = μ₀NI/√(L² + D²). Students record anode current at a fixed anode voltage while stepping the solenoid current upwards, then plot anode current against magnetic field.
In practice the fall is not a vertical step. Electrons leave the cathode with a spread of speeds, the cathode has a finite thickness and the field is not perfectly uniform along the valve, so the curve bends over gradually. The critical field is usually taken where the curve falls most steeply, or where tangents drawn to the flat and falling parts intersect. Repeating at a second anode voltage checks that Bc² rises in proportion to V, as the formula predicts.
Practical Uses
- Determining e/m of the electron from the cut-off of anode current
- Plotting and interpreting a characteristic curve with a gradual rather than sharp transition
- Calculating the field at the centre of a finite solenoid; the current carrying solenoid helps introduce solenoid fields
- Introducing the physics behind the cavity magnetron used in radar and microwave ovens
Specifications
| Quantity determined | Specific charge e/m of the electron |
| Method | Magnetron (critical field cut-off) method with a short solenoid |
| Main parts (typical) | Cylindrical diode valve, short solenoid, heater and anode supplies, ammeter and milliammeter |
| Level | Undergraduate and engineering physics laboratories |
| Valve type, supplied parts and meters | Confirm at enquiry |
Care & Handling
- Allow the valve filament to warm up and the anode current to settle before the first reading.
- Keep the anode voltage constant throughout a set of readings and note it each time.
- Make all connections with the supplies switched off, and keep the valve in its holder so its pins are not strained.
- Switch off the solenoid between sets of readings so its winding does not overheat.
Why Choose LabEquip
College physics departments set up the magnetron method alongside other e/m experiments so that students meet more than one way of measuring the same constant. LabEquip lists this apparatus in its physics lab equipment range; the magnetic focusing method apparatus is the usual companion experiment. Ask about valve type and supplied meters via our contact page.
Frequently Asked Questions
Why is it called the magnetron method?
The valve has a central cathode, a cylindrical anode and a magnetic field along the axis. That is the basic arrangement of a magnetron, the device that generates microwaves in radar sets and microwave ovens.
What is the critical magnetic field?
It is the field at which electrons leaving the cathode curve round just enough to miss the anode. Above it, very few electrons reach the anode and the anode current falls to a small value.
Why does the anode current not drop sharply at one value?
Electrons leave the cathode with a range of speeds, the cathode has some thickness and the field varies slightly along the valve. Different electrons therefore cut off at slightly different fields, which rounds the curve.
How is the field of a short solenoid calculated?
At its centre, B equals the permeability of free space times the number of turns times the current, divided by the square root of the length squared plus the diameter squared. For a long solenoid this reduces to the familiar turns-per-metre formula.
What graph should students plot?
Anode current against the magnetic field, or against solenoid current, at a fixed anode voltage. The critical field is read from the steepest part of the fall, and the value is used with the anode voltage and radius to find e/m.
How can the result be checked?
Repeat the experiment at a second anode voltage. The critical field squared should change in proportion to the voltage, so the calculated e/m should agree within the experimental uncertainty.